Innomatec - S-602-R-2-2315-0.5-16
Leak test computer LTC-602 R Twin functionally identical to device no. 11058 and
9013
Order number: S-602-R-2-2315-0.5-16
Customs tariff number: 90262020
Two-channel version for independent measurement of two test items, with
separate pressure control and start pulse processing, thus is staggered
Measuring and testing with different pressures and limits possible.
Channel I:
Test pressure range: 0.5 - 16 bar relative (electronically controlled)
Pressure difference method relative (R)
Resolution 1/600,000 of the measuring range
Channel II:
Test pressure range: 0.5 - 16 bar relative (electronically controlled)
Pressure difference method relative (R)
Resolution 1/600,000 of the measuring range
Aluminum desktop housing, 560 x 170 x 300 mm (W x H x D),
19" / 3 HU, IP 40 (standard version)
one button operation
Simple and clear menu navigation with Windows technology
5.5" daylight-capable TFT color display with full graphics capability
The test phases are shown on the display
Measurement curve display with pressure and leak rate history
Zoom function for test print curve
Data storage of the last 2000 measurement results in the ring buffer
Data exchange and storage via memory stick, e.g. B. Measurement data acquisition
Electronic precision test pressure regulator
Filling pressure and test pressure (shock filling function)
100 test programs can be programmed and saved
Parameter transfer and program updates with Memory Stick
All limits in ccm³/min, ml/min, l/min, l/h, mbarl/s or Pa, mmWs, mmHG, mbar,
cash freely selectable
All measurement results in ccm³/min, ml/min, l/min, l/h, mbarl/s or Pa, mmWs,
mmHG, mbar, bar freely selectable
RS 232 interface for SPS or for PC
Centronics and USB interface for printers
USB interfaces, front and rear
Ethernet interface for connection to the company network
12 digital potential-free inputs/outputs for external control of a device
Program link depending on an i. Oh message
(NOK message optionally available