(Optional) Dakks Calibration
Isotech
DAkkS calibration of a standard thermometer in the range -38 ° C to 660 ° C
DAkkS calibration of a working standard thermometer to the
Fixed points of the ITS-90. Achievable measurement uncertainties:
Mercury -38.8344 ° C 3.5mK
Water trip point 0.01 ° C 2.5mK
Gallium 29.7646 ° C 2.5mK
Tin 231.928 ° C 7.0mK
Zinc 419.527 ° C 12mK
Aluminum 660.323 ° C 20mK
The achievable measurement uncertainties depend on the
DUT. Some typical measurement uncertainties can be found at:
www.kalibrierdienst-klasmeier.de/mu
The fixed points are determined according to the temperature ranges of the ITS
90 selected. This will be done before the calibration
Working standard thermometer sufficiently aged.
Hydrostatic pressure effects in the fixed-point cells