84DKD-84DX-6050
Burster
Calibration certificate with accreditation symbol (DAkkS)
according to DIN EN ISO 376 for force sensors or
Measuring chains in pressure direction.
10 force levels evenly across the measuring range
distributed (10% increments).
Increasing and decreasing measuring force in different
installation positions.
For measuring range end values 100 N to 50 kN
If the calibration certificate contains a statement of conformity
to comply with error limits, so
we make the determination only under
Consideration of the expanded measurement uncertainty
(confidence level 95%), according to ILAC-G8:09/2019,
2019 Appendix B Example 2.
This means that conformity with
Data sheet information of the calibration object only
can be confirmed provided that the measurement deviation
plus the amount of the extended
Measurement uncertainty within the permissible
specification.
Conformity statements for device classes that are normative
are regulated (DIN 51309 and DIN EN ISO 376)